For optimal performance and accurate readings, TDS meters should be recalibrated periodically using certified calibration solutions. Proper calibration practices help maintain measurement accuracy and extend the life of the meter.
- NaCl calibration solutions are available in 342 ppm or 1000 ppm concentrations.
- Calibration solutions are supplied in lightweight HDPE plastic bottles.
- Never calibrate a meter in air.
- Never calibrate a meter using distilled water.
- Never calibrate a meter using tap water or any uncertified water or solution.
- Unopened HM Digital calibration solutions have a shelf life of approximately one year.
- Once opened, calibration solutions have a shelf life of approximately six months if properly stored and uncontaminated.
- All calibration solutions should be stored in a cool location.